9
Specications and Environmental Conditions
Environmental Conditions
Specications and Environmental Conditions2
Specications2.1
SpecificationsTable 2
Range –10°C to 122°C (14°F to 252°F) at an ambient of 23°C
Accuracy ±0.25°C
Stability ±0.05°C
Resolution 0.1°C or °F
Well-to-Well
Uniformity
±0.2°C with sensors of similar size at equal depths within wells
Heating Times ambient to 100°C: 10 minutes
Stabilization 7 minutes
Cooling Times ambient to 0°C: 10 minutes
Well Depth 4 inches (102 mm)
Removable Inserts refer to Section , Accessories
Power 94–234 VAC (±10%), 50/60 Hz, 50 W; or 12 VDC
Size 4” H x 6” W x 6.9”D
(100 mm x 152 mm x 175 mm)
Weight 4 lb. (1.8 Kg)
Safety Conforms to EN61010-1
Conforms to CAN/CSA C22.2 No.1010.1
UL3111 and ANSI/ISA-S82.01
Fault Protection Sensor burnout protection, over-temperature cutout, and electrical fuses
Fuse Rating 250 V, 3 A FF (very fast acting) NO USER SERVICEABLE PARTS
Environmental Conditions2.2
Although the instrument has been designed for optimum durability and trouble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and cleaning recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5–50°C (41–122°F)
●●
ambient relative humidity: 15–50%
●●
pressure: 75kPa–106kPa
●●
mains voltage within ±10% of nominal
●●
vibrations in the calibration environment should be minimized
●●
altitudes less than 2000 meters
●●
indoor use only
●●
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com