Controller Operation
Serial Interface Parameters
parameters, serial interface parameters, and calibration parameters. The menus are
selected using the up and down arrows and then pressing “SET”.
6.9 Operating Parameters
The operating parameters menu is indicated by,
PAr Operatingparametersmenu
The operating parameters menu contains the High Limit parameter.
6.10 High Limit
The High Limit parameter adjusts the upper set-point temperature. The factory default
and maximum are set to 375°C (707°F). The minimum setting is 125°C (257°F). For
safety, a user can adjust the High Limit down so the maximum temperature set-point is
restricted.
HL FlashesHLandthendisplaysthesetting
375 Flashesthecurrentvalueandthendisplaysthevaluefor
adjustment
375 CurrentHighLimitsetting
Press the u or d to adjust the setting.
300 NewHighLimitsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewHighLimitsetting
6.11 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErL SerialRS-232interfaceparametersmenu
Press “SET” to enter the menu. The serial interface parameters menu contains
parameters which determine the operation of the serial interface. The parameters in the
menu are — baud rate, sample period, duplex mode, and linefeed.
6.11.1 Baud Rate
The baud rate is the rst parameter in the menu. The baud rate setting determines the
serial communications transmission rate.
The baud rate parameter is indicated by,