FLUKE 9102S/725-KIT Lo-Temp Dry-Well and Process Calibrator Kit

User Manual for the Fluke 9102S/725-KIT Lo-Temp Dry-Well and Process Calibrator Kit - Page 36

For 9102S/725-KIT.

PDF File Manual, 165 pages, Read Online | Download pdf file

9102S/725-KIT photo
Loading ...
Loading ...
Loading ...
The calibration parameters are accessed by pressing “SET” after the name of the pa-
rameter is displayed. The value of the parameter may be changed using the up or down
arrow. After the desired value is reached, press “SET” to set the parameter to the new
value. Pressing “EXIT” causes the parameter to be skipped ignoring any changes that
may have been made.
R06.12.1
This probe parameter refers to the resistance of the control probe at 0°C. The value of
this parameter is set at the factory for best instrument accuracy. The value ranges from
95 to 105. For values greater than 100.000, the display does not show the hundreds
placement. For values less than 100.000, the display shows the entire value. The R0
parameter is indicated by,
r0 FlashesforonesecondandthentheR0settingisdisplayed
00.014 Current R0 setting (100.014)
To change the R0 setting, press the up or down arrows.
99.999 New R0 setting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewR0setting
ALPHA6.12.2
This probe parameter refers to the average sensitivity of the probe between 0 and
100°C. The value of this parameter is set at the factory for best instrument accuracy.
aLpha FlashesforonesecondandthentheALPHAsettingis
displayed
38530 CurrentALPHAsetting
To change the ALPHA setting, press the up or down arrows.
38600 NewALPHAsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewALPHAsetting