23
Controller Operation
Heater Power
changed. The set-point resistance is used to perform a calibration adjustment using
the Callendar-Van Dusen R versus T curve t (see Section , Calibration Procedure).
The instrument must be at temperature and stable prior to taking the set-point resis-
tance reading. The set-point resistance can be displayed by pressing the “SET” and d
simultaneously. The set-point resistance is displayed as follows.
SrES SrESashesfortwosecondsandthenthewholenumber
ofthecurrentset-pointresistancesettingisdisplayed
99. Wholenumberportionoftheset-pointresistanceashes
fortwosecondsandthenthefractionportionofthecurrent
set-point resistance setting is displayed
.222 Fractionportionofthecurrentset-pointresistancesetting
The set-point resistance is 99.222.
Secondary Menu6.5
Functions which are used less often are accessed within the secondary menu. The
secondary menu is accessed by pressing “SET” and “EXIT” simultaneously and then
releasing. The rst function in the secondary menu is the heater power display. (See
Figure .)
Heater Power6.6
The temperature controller controls the temperature of the well by pulsing the heater
on and off. The total power being applied to the heater is determined by the duty
cycle or the ratio of heater on time to the pulse cycle time. By knowing the amount of
heating, the user can tell if the calibrator is heating up to the set-point, cooling down,
or controlling at a constant temperature. Monitoring the percent heater power, allows
the user to know the stability of the well temperature. With good control stability the
percent heating power should not uctuate more than ±5% within one minute.
The heater power display is accessed in the secondary menu. Press “SET” and “EXIT”
simultaneously and release. The heater power is displayed as a percentage of full
power.
100.0 C Well temperature
S+E
Accessheaterpowerinsecondarymenu
SEC FlashesSECforsecondarymenuandthendisplaysthe
heaterpower
12.0P Heater power in percent
To exit out of the secondary menu press “EXIT”. To continue on to the proportional
band setting function press “SET”.