9102S Dry-Well Calibrator
Proportional Band
24
Proportional Band6.7
In a proportional controller such as this, the heater output power is proportional to the
well temperature over a limited range of temperatures around the set-point. This range
of temperature is called the proportional band. At the bottom of the proportional band,
the heater output is 100%. At the top of the proportional band, the heater output is 0.
Thus, as the temperature rises the heater power is reduced, which consequently tends
to lower the temperature back down. In this way the temperature is maintained at a
constant level.
The temperature stability of the well and response time depend on the width of the
proportional band. If the band is too wide, the well temperature deviates excessively
from the set-point due to varying external conditions. This is because the power output
changes very little with temperature and the controller cannot respond very well to
changing conditions or noise in the system. If the proportional band is too narrow, the
temperature may swing back and forth because the controller overreacts to tempera-
ture variations. For best control stability, the proportional band must be set for the
optimum width.
The proportional band width is set at the factory as printed on the Report of Calibra-
tion. The proportional band width may be altered by the user if desired to optimize the
control characteristics for a particular application.
The proportional band width is easily adjusted from the front panel. The width may
be set to discrete values in degrees C or F depending on the selected units. The pro-
portional band adjustment is accessed within the secondary menu. Press “SET” and
“EXIT” to enter the secondary menu and show the heater power. Then press “SET” to
access the proportional band.
S+E
Accessheaterpowerinsecondarymenu
SEC FlashesSECforsecondarymenuandthendisplaysthe
heaterpower
12.0P Heater power in percent
S
Accessproportionalband
PrOP Flashespropandthendisplaysthesetting
4.1 Proportionalbandsetting
To change the proportional band press u or d.
10.0 Newproportionalbandsetting
To accept the new setting press “SET”. Press “EXIT” to continue without storing the
new value.
S
Acceptthenewproportionalbandsetting