9102S Dry-Well Calibrator
Scan
22
Scan6.3
The scan rate can be set and enabled so that when the set-point is changed the dry-well
heats or cools at a specied rate (degrees per minute) until it reaches the new set-point.
With the scan disabled the dry-well heats or cools at the maximum possible rate.
Scan Control6.3.1
The scan is controlled with the scan on/off function that appears in the main menu
after the temperature scale units.
Sc Scashesforonesecondandthenthecurrentscansetting
is displayed
Off Scanfunctionoff
Press the up or down arrow to toggle the scan on or off.
On Scanfunctionon
Press “SET” to accept the present setting and continue. Press “EXIT” to cancel.
S
Accept scan setting
Scan Rate6.3.2
The scan rate can be set from 0.1 to 99.9°C/min. The maximum scan rate, however,
is actually limited by the natural heating or cooling rate of the instrument. This rate is
often less than 100°C/min, especially when cooling.
The scan rate function appears in the main menu after the scan control function. The
scan rate units are in degrees Celsius per minute, regardless of the selected units.
Sr Srashesforonesecondandthenthecurrentscanrate
setting is displayed
0.1 Scan rate in°C/min
Press the up or down arrow to change the scan rate.
2.0 New scan rate
Press “SET” to accept the new scan rate and continue. Press “EXIT” to cancel.
S
Accept scan rate
Set-point Resistance6.4
The set-point resistance is the resistance the instrument is trying to make the control
sensor achieve and is calculated in the rmware using the set-point temperature. This
value is not directly adjustable but is recalculated when the set-point temperature is