Chapter 2 To Set the Vertical System RIGOL
DS8000-R User Guide 2-9
integrity of the signal under test. Click Bias, and use the pop-up numeric keypad
or scroll with the mouse to set the DC bias voltage.
âš« Probe Information
Click About to view information about the probe currently connected, such as
the manufacturer, model, serial number, and the last calibration date.
âš« Probe Calibration
Click Calibration to start zero calibration automatically for the probe.
âš« Demagnetization
Click Demagnetize to perform demagnetization automatically for the probe.
Note: This function is only available when the specified probe is connected. For
details about the probe demagnetization operation, refer to
PCA1030/PCA2030/PCA1150 User Guide
.
Amplitude Unit
Select the amplitude display unit for the current channel. The available units are W, A,
V, and U. When the unit is changed, the unit related to the relevant functions of the
channel will change accordingly.
Click the CH1 channel status label to open the CH1 setting menu. Click More → Unit
to select a unit. The default unit for the value is V.
Channel Delay
When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring relatively greater errors (zero offset). DS8000-R allows you to
set a delay time for calibrating the zero offset of the corresponding channel. Zero
offset is defined as the offset of the crossing point of the waveform and trigger level
line relative to the trigger position, as shown in Figure 2-2.