Chapter 19 System Utility Function Setting RIGOL
DS8000-R User Guide 19-11
Figure 19-3 Self-calibration Interface
âš« Click Start, and then the oscilloscope will start to execute the self-calibration
program.
âš« After the self-calibration program is launched, the Exit menu does not function.
You can double-click any place on the screen to stop the self-calibration.
âš« Click Window continuously to open or close the self-calibration information
window. You can also click on the icon
at the upper-right corner of the
self-calibration information window to close the window.
Note: After the self-calibration starts, the "Calibration" window is displayed. The
status shows "Waiting", indicating that the instrument is calibrating. At this time, to
cancel the calibration, double-click any place on the screen. Then, the calibration
status is changed to "Abort". Click Start again, and the oscilloscope starts to
calibrate again. After the calibration has been completed, the status shows "Stop"
again. During the self-calibration, most of the menus are disabled.
Auto Config
Click Utility → More → Auto Config, and set the parameters for the Auto key at
the right section of the Web Control page. For detailed settings, refer to descriptions
in "Quick Measurement after AUTO".
Print Setting
Connect the oscilloscope to the local area network via the network cable. The