RIGOL Chapter 1 Quick Start
1-8 DS8000-R User Guide
voltage adjustment if necessary. For detailed procedures, refer to the "Active
Probe" section.
Function Inspection
This section takes RP3500A passive high-impedance probe as an example to
illustrate the steps of function inspection for the oscilloscope.
1. Click on the function navigation icon
at the lower-left corner of the screen
to open the function navigation. Then click on the "Storage" icon to enter
the storage and loading setting menu. Click More → Default, then a prompt
message "Restore default settings?" is displayed. Click OK to restore the
instrument to its factory default settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in Figure 1-8 below.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe, as shown in Figure 1-8.
Figure 1-8 To Use the Compensation Signal
4. Set the probe attenuation to 10X, and then click the Auto key in the Web
Control page.
5. Observe the waveform on the display. In normal condition, the square waveform
as shown in Figure 1-9 should be displayed.
Figure 1-9 Square Waveform Signal
Compensation Signal Output Terminal
Ground Terminal