FLUKE 525B/A0 120V Temperature/Pressure Calibrator

User Manual for the Fluke 525B/A0 120V Temperature/Pressure Calibrator - Page 84

For 525B/A0 120V.

PDF File Manual, 111 pages, Read Online | Download pdf file

525B/A0 120V photo
Loading ...
Loading ...
Loading ...
Remote Commands
Remote Command Listing
5
5-19
TSENS_TYPE?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Temperature Sensor Type query) Returns the temperature sensor type
thermocouple (TC) or Resistance Temperature Detector (RTD) for temperature
measurements.
Responses: TC (Thermocouple)
RTD (Resistance Temperature Detector)
Example: TSENS_TYPE? returns TC
Returns TC when the temperature sensor type is a thermocouple.
*TST?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Self Test command) Initiates self-test and returns a 0 for pass or a 1 for fail. If
any faults are detected, they are displayed on screen (Terminal mode) or are
logged into the fault queue where they can be read by the FAULT? query
(Computer mode).
Responses: 0 (pass self test)
1 (fail self test)
Example: * TST? returns 1
Returns 1 when self test fails.
VAL?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Measurement Value command) Returns the last value of the resistance,
temperature, or pressure measurement. The second parameter returned is either the
units of the measurement or an indication that an error occurred.
Responses: 1. (Float) Measured resistance, temperature, or pressure
2. (Character) CEL, FAR, OHM, PSI, INH2O4C,
INH2O20C, CMH2O4C, CMH2O20C, BAR, MBAR, KPA,
INHG, MMHG, KG/CM2
OVER (value is over or under capability),
NONE (not in a measurement mode)
Example: VAL? returns 0.000000E+00,NONE
Returns 0 and NONE when the Calibrator is not in a measurement mode.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...