525B
Users Manual
7-2
Electrical Specifications
DC Voltage Specifications, Output
Ranges
[1]
Absolute Uncertainty,
tcal ± 5 °C
± (ppm of output + µV)
Stability
Resolution
Maximum
Burden
[2]
90 days 1 year
24 hours, ± 1 °C
± (ppm of output + µV)
0 to 100.000 mV 25 3 30 3 5 + 2
1 µV
10 mA
0 to 1.00000 V 25 10 30 10 4 + 10
10 µV
10 mA
0 to 10.0000 V 25 100 30 100 4 + 100
100 µV
10 mA
0 to 100.000 V 25 1 mV 30 1 mV 5 + 1 mV 1 mV 1 mA
TC Output
-10 to 75.000 mV 25 3 30 3 5 + 2
1 µV 10
Notes:
[1] All outputs are positive only.
[2] Remote sensing is not provided. Output resistance is < 1.
Ranges
Noise
Bandwidth 0.1 to 10 Hz
± (ppm of output + µV p-p)
Bandwidth 10 Hz to 10 kHz
(µV rms)
0 to 100.000 mV
1 µV 6 µV
0 to 1.00000 V
10 µV 60 µV
0 to 10.0000 V
100 µV 600 µV
0 to 100.000 V 10 ppm+1 mV 20 mV
DC Current Specifications, Output
Ranges
[1]
Absolute Uncertainty
tcal ± 5 °C ± (ppm of output + mA)
Resolution
Maximum
Compliance
Voltage
Maximum
Inductive Load
90 days 1 year
0 to 100.000 mA 40 1 50 1
1 µA
12 V 100 mH
Note:
[1] All outputs are positive only.
Ranges
Noise
Bandwidth
0.1 to 10 Hz p-p
Bandwidth
10 Hz to 10 kHz rms
0 to 100.000 mA 2000 nA
20 µA