Using Input Mode
Using Input Mode
3
3-3
To measure temperature using a thermocouple:
1. Attach the thermocouple leads to the TC input/output connector as shown in
Figure 3-2. One pin is wider than the other. Do not attempt to force the plug in
the wrong polarization.
Note
If the Calibrator and the TC plug are at different temperatures, wait
three minutes or more for the connector temperature to stabilize
after you plug the miniplug into the connector.
2. Press to select TC mode.
3. If necessary, press to place the Calibrator in TC IN mode.
4. Press to select the desired TC type. You can toggle between ºC and ºF
temperature units by pressing .
INPUT
OUTPUT
HI
VOLTS
mA
RTD
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4W RTD
SET RECALL
LOCAL EXP
AUTOSET
INPUT
OUTPUT
ZERO
RNG LOCK
C / F
CJC SETUP
20V PK
MAX
100V MAX
100mA MAX
20V PK
MAX
20V PK
MAX
0
1
2
3
4
5
6
7
8
9
/
VOLTS
mA
TYPE
UNITS
TC
RTD
SHIFT
ENTER
STBY
OPR
CE
TC
fcn07f.eps
Figure 3-2. Measuring Temperature with a Thermocouple