OUT
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Output command) Sets the output of the Calibrator and establishes a new
reference point for the error mode. To source or measure a temperature, select the
desired sensor and sensor parameters first. (See the TSENS_TYPE, RTD_TYPE,
and TC_TYPE commands.)
Use multipliers, k, m, u with the OUT command, as desired.
Parameters: <value> V Volts dc
<value> A Curre nt dc
<value> OHM Resist ance
<value> CEL Temper ature (Celsius)
<value> FAR Temper ature (Fahrenheit)
Examples: OUT 15.2 V (volts; 15.2 V)
OUT 1.2 mA (current; 1.2 mA)
OUT 5 Ohm (ohms; 5 )
OUT 100 CEL (temperature; 100 °C)
OUT 32 FAR (temper ature; 32°F)
Each example shows a value and unit, e.g., 15.2 V. If a value is entered without a
unit, the value of the existing output is changed, when logically allowed.
OUT?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Output query) Returns the output amplitude of the Calibrator. Multipliers (e.g., K
or M) are not used in the response.
Response: <amplitude value>,<units>
Examples: OUT? returns 1.520000E+01,V
OUT? returns 1.88300E02,A
OUT? returns 1.23000E+00,V
OUT? returns 4.00000E+03,OHM
OUT? returns 1.52000E+01,V
OUT? returns 1.0430E+02,CEL
The respective values for the above examples are:
15.2 V
18.8 3 mA
1.23 V
4 K
15.2 V
104. 3 °C