Remote Commands
Remote Command Listing
5
5-13
RANGELCK?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Range command) Returns the RANGELOCK state 1 for ON or 0 for OFF.
Parameter: (None)
REMOTE
OverlappedIEEE-488 RS-232 Sequential
x x xx
(Remote command) Places the Calibrator into the remote state. This command
duplicates the IEEE-488 REN (Remote Enable) message. When the Calibrator is
in the remote state, and not locked out, only the LOCAL key is active. If the front
panel is locked out, no front panel keys are active. See the LOCKOUT command.
To unlock the front panel, use the LOCAL command, or cycle the Calibrator power
switch.
Parameter: (None)
Example: REMOTE
*RST
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Reset Instrument command) Resets the Calibrator to the power-up state. *RST
holds off execution of subsequent commands until the reset operation is complete.
A reset action evokes the following commands and values:
Command Value
OUT 0 V
PRES_UNIT Last selected
RANGE 0.1 V
RTD_TYPE Last selected
STBY (No output)
TC_REF INT
TC_TYPE Last selected
TSENS_TYPE Last selected
Response: (None)
Example: *RST
Place the Calibrator in a reset condition, evoking the commands and values shown
above.