Loading ...
Loading ...
Loading ...
Specifications
Electrical Specifications
7
7-3
Resistance Specifications, Output
Ranges
[1]
Absolute Uncertainty
tcal ± 5°C, ± Ω
Resolution Allowable Current
[2]
90 days 1 year
5 to 400.00 Ω
0.012 0.015
0.001 Ω
1 to 3 mA
5 to 4.0000 kΩ
0.25 0.3
0.01 Ω 100 µA to 1 mA
Notes:
[1] Continuously variable from 0 to 4 kΩ.
[2] For currents lower than shown, the floor adder increases by Floor
(new)
= Floor
(old)
x I
min
/I
actual
.
For example, a 500 µA stimulus measuring 100 Ω has a floor uncertainty of 0.015 Ω x 1 mA/500 µA = 0.03 Ω.
Resistance Specifications, Input
Ranges
[1]
Absolute Uncertainty
tcal ± 5°C, ± (ppm of output + Ω)
Resolution Stimulus Current
90 days 1 year
5 to 400.00 Ω
20 0.0035 20 0.004
0.001 Ω
1 mA
5 to 4.00000 kΩ
20 0.0035 20 0.004
0.01 Ω
0.1 mA
Note:
[1] 4-wire mode.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...