Remote Commands
Remote Command Listing
5
5-15
RTD_TYPE?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Resistance Temperature Detector Type query) Returns the Resistance
Temperature Detector (RTD) type used for RTD temperature simulations.
Responses: PT385_100 (100-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_200 (200-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_500 (500-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_1000 (1000-ohm RTD, curve α=0.0038 ohms/ohm/°C)
PT392_100 (100-ohm RTD, curve α=0.003926 ohms/ohm/°C)
PTJIS_100 (100-ohm RTD, curve α=0.003916 ohms/ohm/°C)
CU10 (10-ohm RTD, empirical curve)
NI120 (120-ohm RTD, empirical curve)
YSI_400 (YSI thermistor curve)
OHMS_HIGH (4000 ohms range)
OHMS_LOW (400 ohms range)
SPRT (Standard PRT with user defined error
coefficients)
USR_DEF<x> (RTD with user defined coefficients (x from
1 to 5)
Example: RTD_TYPE? returns PTJIS_100
Returns PTJIS_100 when a 100-ohm RTD with curve α=0.3926 ohm/°C is set
as the RTD type.
*SRE
Sequential OverlappedIEEE-488 RS-232
x x xx
(Service Request Enable command) Loads a byte into the Service Request Enable
(SRE) register. (See “Service Request Enable Register (SRE)” in Chapter 4. Since
bit 6 is not used (decimal value 64), the maximum entry is 255 64 = 191.
Parameter: <value> (the decimal equivalent of the SRE byte, 0 to 191)
Example: *SRE 48
Enable bits 4 (MAV), and 5 (ESR).