Remote Commands
Remote Command Listing
5
5-9
Set bit 0 of the Event Status Register to 1 when all pending device operations are
done.
*OPC?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Operations Complete query) Returns a 1 after all pending operations are
complete. This command causes program execution to pause until operations are
complete. (See *WAI.)
Response: 1 (all operations are complete)
Example: *OPC? returns 1
Returns 1 when all pending operations are complete.
OPER
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Operate command) Activates the Calibrator output if it is in standby. This is the
same as pressing the Calibrator front panel key.
Parameter: (None)
Example: OPER
Connect the selected output to the Calibrator front panel terminals. Also indicates
OPR on the display.
OPER?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Operate query) Returns the operate/standby setting.
Response: 1 (Operate)
0 (Standby)
Example: OPER? returns 1
Returns 1 when the Calibrator is in operate.
*OPT?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Options command) Returns a list of the installed hardware and software options.
Responses: <option string>,<option string>,(options list, separated by
commas)
0 (no options are installed)
Example: *OPT? Returns 0
This command is reserved for future use.