525B
Users Manual
5-2
Common Commands (continued)
Command Desc ription
*OPT? Returns a list of the installed hardware and software options.
*RST Resets the state of the instrument to the power-up state. This command
holds off execution of subsequent commands until it is complete.
(Overlapped command.)
*SRE Loads a byte into the Service Request Enable register (SRE).
*SRE? Returns the byte from the Service Request Enable register.
*STB? Returns the status byte.
*TST? Initiates a series of self-tests, then returns a "0" for pass or a "1" for fail. If
any faults are detected, they are logged into the fault queue where they
can be read by the FAULT? query.
*WAI Prevents further remote commands from being executed until all previous
remote commands have been executed.
External Connection Commands
Command Desc ription
PRES_UNIT Sets the pressure display units.
PRES_UNIT? Returns the pressure display units.
RTD_TYPE Sets the Resistance Temperature Detector (RTD) type.
RTD_TYPE? Returns the Resistance Temperature Detector (RTD) type.
TC_REF Sets whether the internal temperature sensor or an external reference
value is used for Thermocouple (TC) outputs and measurements.
TC_REF? Returns the source being used as a reference for thermocouple simulation
and measurements.
TC_TYPE Sets the thermocouple (TC) temperature type.
TC_TYPE? Returns the thermocouple (TC) type.
TSENS_TYPE Sets temperature sensor type.
TSENS_TYPE? Returns the temperature sensor type.