Remote Commands
Remote Command Listing
5
5-17
TC_REF
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Thermocouple Reference command) Sets whether the internal temperature sensor
(INT) or an external reference value (EXT) is used for Thermocouple (TC) outputs
and measurements. The Calibrator retains the TC reference setting until power off
or reset.
Parameters: INT
EXT
Example: TC_REF EXT
Set the thermocouple reference to external.
TC_REF?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Thermocouple Reference query) Returns the source of the temperature being
used as a reference for thermocouple simulation and measurement. The choices are
Internal reference (INT) or External reference (EXT).
Responses: INT
EXT
Example: TC_REF? returns INT
Returns Internal.
TC_TYPE
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Thermocouple Type command) Sets the Thermocouple (TC) temperature sensor
type. The TC type is used when the output is set to a temperature value with the
OUT command and the temperature sensor type is set to TC with the
TSENS_TYPE command. When the thermocouple type is changed while
simulating a temperature output, the temperature is changed to 0 °C. The
Calibrator retains the TC type until power off or reset.
Parameters: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple)
L (L-type thermocouple)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
1.888.610.7664 sales@GlobalTestSupply.com