Remote Operation
Setting up the RS-232 Port for Remote Control
4
4-3
Complete the following procedure to test RS-232 port operation using the
Windows
®
HyperTerminal application (or equivalent).
To test port operation:
1. Select Start->Programs->Accessories->Hyperterminal->Hyper Terminal.
2. Enter 525B for Name:, and select OK.
3. Select PC serial port which 525B is connected to (from drop-down list), for
Connect using: and select OK
4. Select the following setting and then select OK:
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: Xon / Xoff
5. Select File->Properties->Settings->ASCII Setup.
6. Select the following settings and then select OK:
Echo typed characters locally
Append line feeds to incoming line ends
7. Select OK.
8. Enter *IDN? and press Enter.
9. Verify that that the returned string is Fluke,525B,0,<firmware revision>.