3-2
V mA
LOOP
V mA
TC R TD
°
C
°
F
Hz
100%
25%
25%
RECALL
ZERO
MEAS
SOURCE
STORE
SETUP
0%
MULTIFUNCTION CALIBRATOR
725
INPUT
OUTPUT
HI
VOLTS
mA
RTD
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4W RTD
SET RECALL
LOCAL EXP
AUTOSET
INPUT
OUTPUT
ZERO
RNG LOCK
C / F
CJC SETUP
20V PK
MAX
100V MAX
100mA MAX
20V PK
MAX
20V PK
MAX
0
1
2
3
4
5
6
7
8
9
/
VOLTS
mA
TYPE
UNITS
TC
RTD
SHIFT
ENTER
STBY
OPR
CE
fcn04f.eps
Figure 3-1. Measuring Resistance Using a 4-Wire to 2-Wire Connection
Measuring Temperature Using Thermocouples
The Calibrator supports numerous standard thermocouples. See Chapter 7 for
supported thermocouple specifications.
You can use the key to toggle the cold junction reference location (CJC)
between the internal and external cold junction reference locations. External cold
junction reference is 0 °C or 32 °F. This is used primarily when you want to use an
external cold junction for conducting temperature measurements. The
XCJC status
message appears on the display when the Calibrator is in external CJC mode.