Using Output Mode
Using Output Mode
2
2-7
Output DC Voltage
The Calibrator is a fully programmable precision source of DC voltage from 0 V
to 100 V. The Calibrator can only output positive (+) values.
To output DC voltage:
1. Connect the Calibrator to the UUT as shown in Figure 2-3.
2. Press for V out display.
3. Use the numeric keypad to enter the desired output value and press . You
can also adjust the output value by pressing . Press to select a
different digit to modify.
4. Press to activate output.
Note
For safety purposes, the Calibrator resets to standby when output is
set to 30 V or greater.
V mA
LOOP
V mA
TC R TD
°
C
°
F
Hz
100%
25%
25%
RECALL
ZERO
MEAS
SOURCE
STORE
SETUP
0%
MULTIFUNCTION CALIBRATOR
725
INPUT
OUTPUT
HI
VOLTS
mA
RTD
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4W RTD
SET RECALL
LOCAL EXP
AUTOSET
INPUT
OUTPUT
ZERO
RNG LOCK
C / F
CJC SETUP
20V PK
MAX
100V MAX
100mA MAX
20V PK
MAX
20V PK
MAX
0
1
2
3
4
5
6
7
8
9
/
VOLTS
mA
TYPE
UNITS
TC
RTD
SHIFT
ENTER
STBY
OPR
CE
fcn02f.eps
Figure 2-3. Connection for Setting DC Voltage Output