525B
Users Manual
5-4
Status Commands
Command Desc ription
FAULT? Returns the latest error code from the error queue, then removes that error
code from the queue.
Error Code Listing
Error
Number
Message
Class
Description
1 DDE Error queue overflow.
100 EXE OPER or STBY was received when the 525B is in Measure mode or
TC Source mode.
101 CME A non-numeric entry was received in a field that should contain a
numeric entry.
102 EXE The numeric field exceeds 10 characters.
103 CME Invalid units symbol or prefix.
104 EXE An attempt to enter RTD Source mode was made when SPRT is
selected.
105 EXE Entry is above upper limit for the selected output range.
106 EXE Entry is below lower limit for the selected output range.
107 EXE Operate not allowed when >30 V and error is pending.
108 CME A required command parameter was missing.
109 CME An invalid TC_MEAS or RTD_MEAS unit parameter (not CEL or FAR)
was received, or an invalid PRES_UNIT parameter was received.
110 CME An invalid RANGELCK parameter (not ON or OFF) was received.
111 EXE RANGELCK ON was received when the 525B is not in Volts mode.
112 CME An invalid RTD_TYPE parameter was received.
113 CME An invalid TC_REF parameter (not INT or EXT) was received.
114 CME An invalid TSENS_TYPE parameter (not TC or RTD) was received.
116 EXE ZERO_MEAS command error. Not allowed or too many parameters.
117 CME An unrecognizable command was received.