FLUKE 525B/A0 120V Temperature/Pressure Calibrator

User Manual for the Fluke 525B/A0 120V Temperature/Pressure Calibrator - Page 71

For 525B/A0 120V.

PDF File Manual, 111 pages, Read Online | Download pdf file

525B/A0 120V photo
Loading ...
Loading ...
Loading ...
525B
Users Manual
5-6
*CLS
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Clear Status command) Clears the ESR, the error queue, and the RQS bit in the
status byte. This command terminates pending operation complete commands
(*OPC or *OPC?).
Parameter: (None)
Example: *CLS
Clear the ESR, the error queue, and the RQS bit in the status byte.
*ESE
Sequential OverlappedIEEE-488 RS-232
x x xx
(Event Status Enable command) Loads a byte into the Event Status Enable (ESE)
register. (See “Event Status Enable Register (ESE)” in Chapter 4.)
Parameter: <value> (decimal equivalent of the ESE byte, 0 to 255)
Example: *ESE 140
Load decimal 140 (binary 10001100) to enable bits 7 (PON), 3 (DDE) and
2 (QYE).
*ESE?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Event Status Enable query) Returns the contents of the Event Status Enable
(ESE) register. (See “Event Status Enable Register (ESE)” in Chapter 4)
Response: <value> (decimal equivalent of the ESE byte, 0 to 255)
Example: *ESE? returns 133
Returns decimal 133 (binary 10000101) when bits 7 (PON), 2 (QYE), 1 (OPC)
are enabled.