FLUKE 525B/A0 120V Temperature/Pressure Calibrator

User Manual for the Fluke 525B/A0 120V Temperature/Pressure Calibrator - Page 81

For 525B/A0 120V.

PDF File Manual, 111 pages, Read Online | Download pdf file

525B/A0 120V photo
Loading ...
Loading ...
Loading ...
525B
Users Manual
5-16
*SRE?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Service Request Enable query) Returns the byte in the Service Request Enable
(SRE).
Response: <value> (the decimal equivalent of the SRE byte, 0 to 191)
Example: *SRE? returns 48
Returns 48 when bits 4 (MAV), and 5 (ESR) are enabled.
*STB?
Sequential OverlappedIEEE-488 RS-232
x x xx
(Status Byte Register query) Returns the byte for the Status Byte Register. See
“Status Byte Register (STB)” in Chapter 4.
Response: <value> (the decimal equivalent of the STB byte, 0 to 255)
Example: *STB? returns 96
Returns 96 if bits 5 (ESR) and 6 (MSS) are set.
STBY
OverlappedIEEE-488 RS-232 Sequential
xx xx
(Standby command) Deactivates the Calibrator output if it is in operate. This is
the same as pressing the Calibrator front panel key.
Parameter: (None)
Example: STBY
Disconnect the selected output from the Calibrator front panel terminals.
TC_MEAS
Sequential OverlappedIEEE-488 RS-232
x x xx
(Thermocouple Measure command) Selects the measure thermocouple mode.
Parameters: CEL (Celsius) (optional)
FAR (Fahrenheit) (optional)
Example: TC_MEAS CEL
Measure the thermocouple temperature that is attached to the Calibrator TC
terminals, in Celsius.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...