SDS2000X HD Series Digital Oscilloscope User Manual
9 2 i n t . s i g l e n t . c o m
input signal is greater than the specified lower limit of time and lower than the specified upper limit of
time value.
Outside a range of time value
--][---- Trigger when the positive or negative slope time of the input
signal is greater than the specified upper limit of time and lower than the specified lower limit of time
value.
Holdoff, coupling, and noise reject can be set in slope trigger, see the sections "Holdoff", "Trigger
Coupling" and "Noise Reject" for details.
15.5.4 Pulse Trigger
Trigger on a positive or negative pulse with a specified width. Trigger source, polarity (positive,
negative), limit range, and time value can be set in the trigger dialog box.
Less than a time value
-- Trigger when the positive or negative pulse time of the input signal is
lower than the specified time value. Below is an example of a trigger condition set to a positive pulse
width < 100ns
Great than a time value
-- Trigger when the positive or negative pulse time of the input signal
is greater than the specified time value. Below is an example of a trigger condition set to a positive
pulse width > 100ns