SDS2000X HD Series Digital Oscilloscope User Manual
i n t . s i g l e n t . c o m 2 6 7
There are 6 tabs in the window. Under each tab, a catalog of waveforms is listed. Built-in waveforms
are stored in Common, Math, Engine, Window, and Trigo. Stored waveforms are located in the Stored
menu.
Users can edit arbitrary waveforms using SIGLENT EasyWaveX PC software, send the stored
waveforms to the instrument through the remote interface, or import the stored waveforms through a
U disk.
Other Setting
Touch Setting in the AWG dialog box to recall the "Other
Setting" dialog box:
A. Select the output load
B. Turn on/off the OVP (Over Voltage Protection)
C. Turn on/off the sync output. Only supported for
external SAG1021I function/arbitrary waveform
generator module.
D. Do zero adjust
E. Set to default
F. Return to the previous menu
Output Load
The selected output load value must match the load impedance. Otherwise, the amplitude and offset
of the output waveform of AWG will be incorrect.
OVP
When OVP is enabled, the output will be turned off automatically once the protection condition is met.
The protection condition is when the absolute value on the output port is higher than 4 V ± 0.5 V. At
the same time, a warning message is displayed.