Measurement Statistics
Measurement statistics are based on the total number of captured waveforms. In Roll mode,
measurement statistics increase over time. Touch Statistics Setting in the measure dialog box to
recall the Statistics Config dialog box:
A. Set the maximum number of samples for the statistics
function. The setting range is 0 ~ 1024, or unlimited. If there
is no limit, the number of statistics will be accumulated. If there
is a limit, when the maximum number of statistics N is
reached, only the latest N measurements will be counted.
When the count is limited, only the first period measurement
value of each frame is counted
B. Set the upper limit of the statistic number in a frame. Valid only
when the count limit is set to "Unlimited". The setting range is
1 ~ 25,000
C. Clear and restart the statistics. Pressing the Clear Sweeps
button on the front panel has an equivalent effect
D. Turn the histogram on or off
E. Return to the previous menu
Enable the Statistics function to observe the distribution of the measured values of every selected
parameter.
Value The current measurement
Mean Average of all historical measurements
Min The minimum of all historical measurements
Max The maximum of all historical measurements
Stdev Standard deviation of all historical measurements
Count The number of historical measurements