SDS2000X HD Series Digital Oscilloscope User Manual
1 2 2 i n t . s i g l e n t . c o m
The data value is displayed in white.
A red point at the end of a segment indicates there is not enough space on the display to
show the complete content of a frame, and some content is hidden, such as .
List
TIME The horizontal offset of the current data frame head relative to the trigger position.
Address Address value. For example, "0x2AB" means that address = 2AB with
acknowledgement.
R/W Read address or write address.
DATA Data bytes.
Configuration
There is only one item Include R/W Bit in the configuration of the I2C decode. When it is disabled,
the address is represented separately from the R/W bit, and when it is enabled, the R/W bit is
represented together with the address.
For example, the address 0x4E: Write: Ack, is displayed as "0x4E (W)" when the R/W bit is not
included and is displayed as "0x9C" when the R/W bit is included.