Siglent SDS2204X-HD 200MHz 4ch Digital Oscilloscope

User Manual - Page 101

For SDS2204X-HD.

PDF File Manual, 299 pages, Read Online | Download pdf file

SDS2204X-HD photo
Loading ...
Loading ...
Loading ...
SDS2000X HD Series Digital Oscilloscope User Manual
i n t . s i g l e n t . c o m 9 9
Trigger source, slope (rising, falling), limit range, and time value can be set in the trigger dialog box.
Holdoff, coupling, and noise reject can be set in interval trigger, see the sections "Holdoff", "Trigger
Coupling" and "Noise Reject" for details.
15.5.8 Dropout Trigger
Dropout trigger includes two types: Edge and State.
Edge
Trigger when the time interval (T) from when the rising edge (or falling edge) of the input signal
passes through the trigger level to when the neighboring rising edge (or falling edge) passes through
the trigger level is greater than the set time, as shown in the figure below:
State
Trigger when the time interval (T) from when the rising edge (or falling edge) of the input signal
passes through the trigger level to when the neighboring falling edge (or rising edge) passes through
the trigger level is greater than the set time, as shown in the figure below:
Trigger source, slope (rising, falling), dropout type, and time value can be set in the trigger dialog box.
Holdoff, coupling, and noise reject can be set in the dropout trigger, see the sections "Holdoff",
"Trigger Coupling" and "Noise Reject" for details.