1 1 8 i n t . s i g l e n t . c o m
Restart The oscilloscope will be triggered when another “Start” occurs before a Stop".
No Ack The oscilloscope will be triggered when the SDA line is high during any SCL’s ACK bit.
EEPROM The trigger searches for EEPROM control
byte value 1010xxx on the SDA bus. And there is a Read
bit and an ACK bit behind EEPROM. Set the data value and
compare type according to Data1 and Limit Range .
If EEPROM’s data is greater than (less than, equal to)
Data1 , the oscilloscope will be triggered at the edge of
the ACK bit behind the Data byte. The Data byte doesn’t
need to follow the EEPROM.
7 Address & Data the oscilloscope will be triggered on the read or write frame in 7-bit address
mode.
Frame (Start: 7-bit address: R/W: Ack) Data1 and Data2 are set to “0xXX”. If all bits match, then
trigger on the Ack bit followed by the R/W bit.