SDS2000X HD Series Digital Oscilloscope User Manual
i n t . s i g l e n t . c o m 3 7
3. Press the Auto Setup button.
4. Check the waveform displayed and compare it with the following.
Under
Compensated
Perfectly
Compensated
Over
Compensated
5. Use a non-metallic driver to adjust the low-frequency compensation adjustment hole on the
probe until the waveform displayed is as the “Perfectly compensated” in the figure above.
It’s not necessary to compensate active probes like the SIGLENT SAP series.
6.3.7 Logic Probe
To connect the logic probe: Insert the probe, with the correct side facing up, until you hear a “click”.
To remove the logic probe: Depress the buttons on each side of the probe, then pull out it.
See chapter “Digital Channels” for more information.