SDS2000X HD Series Digital Oscilloscope User Manual
6 0 i n t . s i g l e n t . c o m
Bandwidth Limit
Full bandwidth passes through signals with high-frequency components, but it also means that noise
with high-frequency components can pass through. When the frequency component of the signal is
very low, better signal-to-noise ratios (SNR) can be obtained by turning on a bandwidth limit. The
SDS2000X HD provides two hardware bandwidth limit options: 20 MHz and 200 MHz.
Probe Settings
The SDS2000X HD provides 1X, 10X, 100X, and custom probe attenuation factor options. The
custom values can be between 10
-6
~10
6
. The oscilloscope will automatically convert the vertical scale
according to the current probe attenuation factor. For example, the vertical scale of the oscilloscope
under 1X attenuation is 100 mV/div, and the vertical scale will be automatically set to 1 V/div if the
probe attenuation is changed to 10X. If a standard probe with a readout terminal is connected, the
oscilloscope will automatically set the probe attenuation to match the probe.
The Probe Check option is provided to quickly check if the probe is correctly compensated. Connect
the probe as in section “Probes” and touch Probe
Probe Check , then operate according to
the prompt.
Label
Users can set the label text for channels. Open the label setting dialog box to select the source,
customize the label text, and set the display. The source can be C1~C4, Math, and RefA~RefD. The
length of the label is limited to 20 characters. The characters beyond this length will not be displayed.
When the “Display” option is set to “on”, the label will be displayed on the right side of the channel
offset indicator.
The labels are hidden The labels are displayed