26 Histogram
Overview
The SDS2000X HD supports waveform histograms for observing probability distributions of the
waveform in the specified region. The statistics can be performed in both horizontal and vertical
directions. The histogram continues to update as long as the acquisition is active.
Touch Analysis > Histogram to recall Histogram dialog box:
A. Turn on or off the histogram
B. Select the source of the histogram: C1~C4. When
Zoom is turned on, the source selection is
automatically switched to Z1-Z4 (Z = Zoom waveform)
C. Set the type of histogram: Horizontal, Vertical or Both
D. Set the region of the histogram. The setting range is
within the waveform area.
E. Turn on or off histogram statistics
F. Clear and restart the histogram statistics. Pressing the
Clear Sweeps button on the front panel has an
equivalent effect
Type
Horizontal Displays the histogram in the horizontal direction. The oscilloscope counts waveform
data falling into every horizontal (time) bin defined by the histogram region and shows the
histogram at the bottom of the waveform area.
Vertical Displays histogram in the vertical direction. The oscilloscope counts waveform data
falling into every vertical (amplitude) bin defined by the histogram region and shows the histogram
at the left of the waveform area.
Both Displays both horizontal and vertical histograms