Loading ...
Loading ...
Loading ...
SDS2000X HD Series Digital Oscilloscope User Manual
i n t . s i g l e n t . c o m 1 0 3
When the type is “Edge”, the oscilloscope triggers at the first edge after the specified edge (Rising or
Falling) of the qualified source; when the type is “Edge with Delay”, a time limit condition is available.
Qualified
Source
Edge
Trigger
Source
Trigger
Position
Qualified Edge = Rising
Edge = Rising
Touch the Qualified Setting region to set the qualified source and threshold;
Touch the Edge Setting region to set the edge trigger source, threshold, and slope.
15.5.12 Nth Edge Trigger
Nth edge trigger is similar to an edge trigger with conditions. The trigger is satisfied after meeting the
user-defined idle time and edge number conditions. As shown in the figure below, when the idle time
between the pulse trains is greater than the specified idle time, it is triggered on the third falling edge
of the pulse train.
Trigger source, slope (rising, falling), idle time, and edge number can be set in the trigger dialog box.
Holdoff, coupling, and noise reject can be set in the Nth edge trigger, see the sections "Holdoff",
"Trigger Coupling" and "Noise Reject" for details.
Loading ...
Loading ...
Loading ...