Siglent SDS2204X-HD 200MHz 4ch Digital Oscilloscope

User Manual - Page 149

For SDS2204X-HD.

PDF File Manual, 299 pages, Read Online | Download pdf file

SDS2204X-HD photo
Loading ...
Loading ...
Loading ...
SDS2000X HD Series Digital Oscilloscope User Manual
i n t . s i g l e n t . c o m 1 4 7
If you set the trigger condition to Slow Channel:
ID can be selected in the hexadecimal range of 0x0 to
0xF (short serial / enhanced serial with 4 bits IDor 0x00 to
0xFFenhanced serial with 8 bits ID. If it is selected as
"0xXX", the ID is ignored.
Compare Type can be specified as Equal, Not Equal,
Less than, Less or Equal, Greater than, Greater or Equal,
in Range of, out of Range.
Data can be selected in the hexadecimal range of 0x00
to 0xFF (short serialor 0x0000 to 0xFFFFenhanced serial
with 4 bits IDor 0x000 to 0Xfff(enhanced serial with 8 bits
ID). If it is selected as "0xXX", the data is ignored.
Error the oscilloscope triggers on the error frame. Errors include Successive Sync Pulses Error
Pulse Period Error, Fast Channel CRC Error, Slow Channel CRC Error, All CRC Errors.
Successive Sync Pulses Error: Triggers on a sync pulse whose width varies from the previous
sync pulse's width by greater than 1/64 (1.5625%, as defined in the SENT specification).
Pulse Period Error: Triggers if a nibble is either too wide or too narrow (for example, data nibble
< 12 (11.5) or > 27 (27.5) ticks wide). Sync, S&C, data, or checksum pulse periods are checked.
Fast Channel CRC Error: Triggers on any Fast Channel Message CRC error.
Slow Channel CRC Error: Triggers on any Slow Channel Message CRC error.
All CRC Errors: Triggers on any CRC error, Fast or Slow.