SDS2000X HD Series Digital Oscilloscope User Manual
3 6 i n t . s i g l e n t . c o m
6.3.5 Waveform Generator
The SDS2000X HD has a built-in function/arbitrary waveform generator module installed at the factory.
Purchase the "SDS2000HD-FG" option to activate the function generator function.
You can also purchase an external SAG1021I function/arbitrary waveform generator module with
isolated output to use as a waveform generator. Connect the SAG1021I to any USB host port of the
oscilloscope through a USB cable. To use the SAG1021I, you also need to purchase the
"SDS2000HD-FG" software activation option.
Press the WaveGen button on the front panel or touch the screen Utility
Wave Gen to
set the waveform. See chapters Arbitrary Waveform Generator and Bode Plot for more relative
information.
6.3.6 Probes
The SDS2000X HD series oscilloscope supports active probes and passive probes. The
specifications and probe documents can be obtained at
int.siglent.com, www.siglentna.com, or
www.siglenteu.com.
Probe Compensation
When a passive probe is used for the first time, you should compensate it to match the input channel
of the oscilloscope. Non-compensated or poorly compensated probes may increase measurement
inaccuracy or error. The probe compensation procedures are as follows:
1. Connect the coaxial cable interface (BNC connector) of the passive probe to any channel
of the oscilloscope.
2. Connect the probe to the “Compensation Signal Output Terminal” (Cal) on the front of the
oscilloscope. Connect the ground alligator clip of the probe to the “Ground Terminal” under
the compensation signal output terminal.