Siglent SDS2204X-HD 200MHz 4ch Digital Oscilloscope

User Manual - Page 196

For SDS2204X-HD.

PDF File Manual, 299 pages, Read Online | Download pdf file

SDS2204X-HD photo
Loading ...
Loading ...
Loading ...
SDS2000X HD Series Digital Oscilloscope User Manual
1 9 4 i n t . s i g l e n t . c o m
Split Mode, Zoom off Split Mode, Zoom on
Full-Screen Mode Exclusive Mode
FFT Mode
Normal: Displays the FFT result of each frame directly.
Max-Hold: Holds the maximum value in the historic frame on the display until cleared. This mode
is suitable for detecting discontinuous waves, such as sporadic pulse signals, or frequency
hopping signals. Press the Clear Sweeps button on the front panel to clear the max-hold
waveform.
Average: Reduces the effect of the superimposed random noise on the signal. After the FFT
mode is set as Average, Average will appear under FFT Mode . The average count can
be set here, ranging from 4 to 1024. Press the Clear Sweeps on the front panel or touch
Reset in the dialog box to restart the average counter.