SDS2000X HD Series Digital Oscilloscope User Manual
1 0 4 i n t . s i g l e n t . c o m
15.5.13 Delay Trigger
A delay trigger is true when the edge of source B occurs after meeting the set conditions of source A
and a user-defined delay time. The setting of source A is similar to that of the pattern trigger and can
be used for logical "and" combination of multiple channels.
Touch the SourceB Setting region to set the edge trigger source, threshold, and slope.
Touch the Limit Range and Upper/Lower Value region to set the delay time condition.
15.5.14 Setup/Hold Trigger
Clock source and data source need to be set in the setup/hold trigger setting. The set-up time starts
when the data signal crosses the trigger level and ends when the specified clock edge arrives. The
holding time starts when the specified clock edge arrives and ends when the data signal crosses the
trigger level again (as shown in the figure below). When the set-up time or hold time meets the preset
time limit conditions, the oscilloscope will trigger.