RIGOL Chapter 5 To Trigger the Oscilloscope
5-32 MSO7000/DS7000 User Guide
Nth Edge Trigger
Triggers on the Nth edge that appears after the specified idle time. For example, in
the waveform shown in the figure below, the instrument should trigger on the
second rising edge after the specified idle time (the time between two neighboring
rising edges), and the idle time should be within the range between P and M (P<Idle
Time<M). Wherein, M is the time between the first rising edge and its previous rising
edge; P is the maximum time between the rising edges that participate in counting.
M
P<Idle Time<M
Edge Type=
Edge Num=2
P
Figure 5-12 Nth Edge Trigger
Trigger Type:
Press Type, and then rotate the multifunction knob to select "Nth Edge". Press
down the knob to select the trigger type. Then, the current trigger setting
information is displayed at the upper-right corner of the screen, as shown in the
figure below. You can also press Type continuously to select the trigger type or
enable the touch screen to tap the desired trigger type and select it.
Source Selection:
Press Source to open the signal source list and select CH1-CH4 or D0-D15. For
details, refer to descriptions in "Trigger Source". The current trigger source is
displayed at the upper-right corner of the screen.
Note: Only when we select the channel (that has been input with signals) as the
trigger source, can we obtain a stable trigger.
Edge Type:
Press Slope continuously to select the input signal edge on which the oscilloscope
triggers.
âš«
Rising: triggers on the rising edge of the input signal when the voltage level
meets the specified trigger level.
âš«
Falling: triggers on the falling edge of the input signal when the voltage level
meets the specified trigger level.