CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 74/197
7.3 Special controller functions via operation lines
Press the Setpoint setting icon to access the “Setpoint” setting menu from Normal display.
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Chamber with O
2
control: Depending on the required range of O
2
concentration, the type of gas which is
required may vary (chap. 6.6). Therefore, the pressure alarms for O
2
and N
2
can be deactivated individually
with active O
2
control.
For chambers equipped with the optional interior socket (chap. 18.4) you can turn on and off the voltage of
the interior socket via the controller.
• Operation line “Idle mode” activates / deactivates the operating mode “Idle mode”.
• Operation line “O2 pressure alarm off” activates / deactivates the O
2
pressure alarm (chamber with O
2
control)
• Operation line “N2 pressure alarm off” activates / deactivates the N
2
pressure alarm (chamber with O
2
control)
• Operation line “Interior socket” activates / deactivates interior socket (option, available via BINDER
INDIVIDUAL Customized Solutions)
The other operation lines are without function.
Use the “Setpoints” menu to configure the operation lines.
“Setpoints” menu.
Select the field “Functions on/off”.
“Functions on/off” entry menu (view with
chamber with O
2
control with optional interior
socket).
Mark / unmark the checkbox to activate /
deactivate the desired function and press the
Confirm icon
Activated operation line: switching status “1” (On)
Deactivated operation line: switching status “0” (Off)
The operation lines count from right to left.
Example:
Activated operation line “Idle mode” = 000000000000000
1
Deactivated operation line “Idle mode” = 000000000000000
0