22.2 CBF / CBF-UL: Replacing components of the humidifying / dehumidifying
system
22.2.1 Replacing the silicone tube of the fresh water supply system
The silicone tube (DN 3mm) (art. no. 6008-0267) is included with the following humidifying sets:
Tubular bag set with 3 tubular bags (Art. No. 8009-1083, included): 6 m (chap. 4.5.2)
Tubular bag set with 12 tubular bags (Art. No. 8009-1058, option): 6 m (chap. 4.5.2)
Humidification water set (Art. No. 8009-1104, option): 2 m (chap. 4.5.3)
Water container set (Art. No. 8009-1128, option): 10 m (chap. 4.5.4)
It is also available by the meter from BINDER.
To operate the freshwater supply via tubular bags (standard), a total length of 32 cm is required.
To replace the DN 3mm silicone tube, proceed as described in chap. 4.5.
22.2.2 Replacing the pump head of the fresh water supply system
A pump head for the tube pump for freshwater supply is available under art.no. 5013-0204. To replace the
pump head, proceed as described in chap. 4.5.
22.2.3 Replacing the gas fine filter (sterile filter) for ambient air
The air tube for inlet of the ambient air is located in the upper right corner in the recess for the freshwater
bag. We recommend replacing the gas fine filter (sterile filter) on this tube about once a month. To do this,
hold the sterile filter and pull out a bit the air tube connected to it. Unscrew the filter and screw on a new
filter. Then push the tube back in.
Figure 40: Location of the gas fine filter (sterile filter) in the recess for the freshwater bag