CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 45/197
to freshwater bag
to incubator connection
Figure 17: Inserting the silicone tube into the pump head ④
• Insert the pump head (12) into the pump connector (13) and turn it clockwise until it snaps into place.
• Attach the silicone tube to the incubator connection ⑤
• Place the freshwater bag with the opening to the left in the recess behind the chamber door.
4.5.2.3 Refilling the already installed freshwater bag
For easier handling, you can temporarily release the pump head from the pump connector
on the incubator.
Standard procedure
• Remove the connection between the both parts of the tube connector①.
• Depending on the diameter of the object used for filling (hose, spray bottle ...), you may remove the
connection between the hose connector part and the freshwater bag ①.
• Fill the freshwater bag. Observe water quality and temperature (chap. 4.5.1.3).
• If it has been removed, put the tube connector part with a 5-7 mm nozzle back on the freshwater bag
①
• Reassemble the two parts of the tube connector ①.
Procedure with optional filling hose
A bag set (Art.no. 8009-1058) incl. accessories (annual demand) is optionally available (chap. 4.5.2). For
easier filling, the set contains a filling tube (DN 10 mm) of 1 m length with a fitting tube connector.
• Remove the connection between both parts of the tube connector ①
• Attach the filling tube to the tube connector part which is still on the freshwater bag
• Fill the freshwater bag. Observe water quality and temperature (chap. 4.5.1.3).
• Remove the filling tube from the tube connector on the freshwater bag
• Reassemble the two parts of the tube connector ①.