CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 165/197
After 8760 operating hours or two years the following message appears:
After confirmation with the Confirm icon, the
message window will pop up again every two
weeks until it is reset by BINDER Service.
22.6 Simple troubleshooting
Defects and shortcomings can compromise the operational safety of the chamber and can lead to risks and
damage to equipment and persons. If there are is a technical fault or shortcoming, take the chamber out of
operation and inform BINDER Service. If you are not sure whether there is a technical fault, proceed
according to the following list. If you cannot clearly identify an error or there is a technical fault, please
contact BINDER Service.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.
22.6.1 General
Fault description
Possible cause
Required measures
Chamber without function.
No power supply.
Check connection to power supply.
Check whether the chamber is turned
on at the main power switch.
Wrong voltage.
Check power supply for voltage of 100-
120V or 200-240V.
Chamber fuse has responded.
Check chamber fuse and replace it if
appropriate. If it responds again,
contact BINDER service.
Controller defective.
Contact BINDER service.
Nominal temperature
exceeded by 10 °C due to
chamber failure. Over
temperature protective device
(class 1) responds.