Binder CB 260 CO₂ incubator with hot air sterilization, series CB – 220-liter capacity

User Manual - Page 48

For CB 260.

PDF File Manual, 197 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 48/197
Please proceed in the following order:
Turn the pump head (12) counterclockwise until it stops and remove it from the pump connector (13).
Pull up the tongue of the tube retainer (12b) up to Clickand remove it. Take the rotor (12c) out of the
pump head cover (12a).
Lead the silicone tube along the rotor. Insert the rotor with the tube into the pump head cover. Keep the
hose tight and do not pinch it.
The end of the tube for connecting to the incubator should be about 6 cm long. With a total length of the
silicone tube of 32 cm, the tube end for connecting the water bottle has a length of approx. 18 cm
Press the tube retainer (12b) into the pump head cover until it snaps into place.
to water bottle
to incubator connection
Figure 19: Inserting the silicone tube into the pump head
Insert the pump head (12) into the pump connector (13) and turn it clockwise until it snaps into place.
Attach the silicone tube to the incubator connection.
Plug the part of the tube connector with 3-5mm nozzle in the silicon tube.
Connect the free nozzle of the tube connector with a part of the silicon tube DN 6 mm
Remove the protective cap of the spike connector (transfer set) and connect the nozzle of the tube
connector with the spike connectors using a part of the DN 6 mm silicone tube.
Silicone tube of the
water system
Two-piece tube
connector
Silicone tube
DN 6 mm
Spike connector
(transfer set)
Then remove the second protective cap of the spike connector and insert the piercing pin into the water
bottle through the silicone seal.