Binder CB 260 CO₂ incubator with hot air sterilization, series CB – 220-liter capacity

User Manual - Page 52

For CB 260.

PDF File Manual, 197 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 52/197
Please proceed in the following order:
Assemble the magnetic can holder (18): Attach the water can support to the holder
Attach the magnetic can holder (18) to the left chamber side
Insert the freshwater can (17) from above into the water can support
Screw the hose coupling with aqua stop (19) on the water can outlet
Insert the angular coupling plug (20) into the hose coupling. The gray removal button (19a) on the hose
coupling snaps into place.
Attach a 5 cm piece of the silicone tube d10 (21) to the angular coupling plug.
Insert the reducing connector (22) into the silicone tube (21)
Secure both sides of the silicone tube (21) each with a wire tension clamp (23)
Water supply for a single chamber:
Plug the silicone tube DN 3mm (24) coming from the pump (12) on the reducing connector (22)
Water supply for several chambers:
You can branch off the water supply via T-pieces for additional chambers.
Divide the silicone tube DN 3mm (24) into pieces of a suitable length. Take into account the height of
the magnetic can holder
Put a piece of the silicone tube DN 3mm (24) on the reducing connector (22) and connect a T-piece to
the free end
If there are more than two chambers, you can connect another T-piece using a piece of the silicone tube
DN 3mm (24)
For each chamber, plug the silicone tube DN 3mm (24) coming from the pump (12) onto the free outlet
of the T-piece
Now you can fill the freshwater can. Observe water quality and temperature. Do not firmly close the lid of
the freshwater can to allow air to enter the freshwater can. As a result, obstruction of the water supply due
to negative pressure inside the can is avoided.
4.5.4.3 Refilling the freshwater can
For refilling later, you can remove the freshwater can from the water can support. Press the grey removal
button on the hose coupling, then the angle coupling plug can be removed. A small amount of water can
escape (filling amount of the angular coupling plug). The water can outlet remains closed due to the aqua
stop.
Alternatively, you can remove the lid and fill the can on site with a suitable aid (water hose, watering can).
When filling on site, the existing hose connection can remain on the freshwater can.