18.4 Interior socket 230V (option, available via BINDER INDIVIDUAL Customized
Solutions)
The interior socket is activated and deactivated in the controller via the switching
state of an operation line (chap. 7.3, 9.7.3, 10.6.5). When the interior socket is
activated, the notification “Interior socket“ is shown highlighted in black next to the
information icon in the screen header in Normal display (chap. 11.1.1)
The interior socket (G) is located at the upper left corner on the rear wall of the inner chamber. It is closed
with a waterproof lid. The interior socket is suitable to supply electrical devices inside the chamber with
230V AC voltage.
Figure 32: Position of the interior socket 230 V
(G) Interior socket
The maximum head load must not exceed a capacity of 20 W.
NOTICE
Danger of damage when exceeding the setpoint temperature.
Damage to the cultures.
Do NOT exceed the maximum head load of 20 W.
Do NOT connect equipment with a nominal capacity > 20 W.
Heat emission of electrical devices connected inside the CO
2
incubator may modify the
temperature range.
The maximum load must not exceed 3 Amp.
DANGER
Electrical hazard through overload of contacts.
Deadly electric shock. Damage to contacts and connection socket.
Do NOT exceed the maximum load of 3 A.
Do NOT connect any devices with a nominal current > 3 A.
(G)