CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 153/197
Recommended precautions: To protect the eyes use sealed protective goggles.
CAUTION
Danger of chemical burns through eye contact with the disinfectant
spray.
Eye damage. Environmental damage.
Do NOT empty the disinfectant spray into drains.
Wear protective goggles.
Following frequent use of the disinfectant spray and prior to hot-air sterilization, remove any
agent remainder by using the neutral cleaning agent and then a moistened towel to avoid
formation of permanent residues.
After using the disinfectant spray, allow the chamber to dry thoroughly, and aerate it
sufficiently.
21.2.1 Disinfection of the CO
2
sensor
To ensure proper disinfection and function of the CO
2
sensor, BINDER recommends wiping the sensor
head with pure alcohol or a non-corrosive, acid and chlorine-free alcohol based surface disinfectant. We
recommend using the disinfectant spray Art. No. 1002-0022.
Important: To maintain the proper function of the CO
2
sensor, never spray the sensor with any cleaning
detergents or disinfectants. Generally perform any disinfection only when the chamber is turned off. The
sensor may be only superficially disinfected with a damp cloth. The hot air sterilization routine is intended
for a thorough sterilization of the entire chamber.
NOTICE
Danger of damage to the CO
2
sensor due to unsuitable disinfection procedures.
Damage to the CO
2
sensor.
NEVER spray the disinfectant directly on the CO
2
sensor.
Wipe the sensor with a lint-free cloth soaked with the disinfectant.
Use only the detergents or disinfectants recommended by BINDER. Other products are not approved, since
they could in particular damage the CO
2
sensor.
21.2.2 CBF / CBF-UL: Disinfection of components of the fresh water supply system
Freshwater is supplied by manually filling a freshwater bag (tubular bag set, standard, can be reordered)
or a freshwater can (optional water container set). It is also possible to connect prefilled water bottles
(optional humidification water set).
The freshwater bag can be refilled several times, but should be replaced from time to time for hygienic
reasons.
The silicone tubes and tube connectors can be autoclaved. They can also be replaced for hygienic reasons.
For the fresh water can we recommend thorough cleaning using a recommended cleaning agent (chap.
21.1) followed by rinsing with 70 % ethanol.