CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 64/197
5.6 Principle of controller entries
In the selection and entry menus there are icons displayed in the footers which you can use to take over
the entry or cancel it.
Selection menu (example)
Entry menu (example)
After completing the settings there are the following possibilities:
Press the Confirm icon to take over the entries and exit the menu or continue the menu
sequence.
Press the Close icon to exit the menu or cancel the menu sequence without taking over the
entries.
When terminating a menu sequence, an information window appears, which must be
confirmed.
5.7 Performance during and after power failures
During a power failure, all controller functions are shut down. The gas inlet valves are closed so that no gas
can escape into the ambient air. The zero-voltage relay alarm output (3) (chap. 11.6) is switched to alarm
position for the whole duration of the power failure.
After the power returns, all functions return to the same status the chamber had before power failure. . The
controller continues to function in the original operating mode it was in previously before the power failure
occurred.
Performance after power failure in Idle mode
Control is deactivated
Performance after power failure in Fixed value operation mode
All functions return to the same status the chamber had before power failure. The set-points are
immediately resumed.
Performance after power failure during time program operation
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run.
Performance after power failure during week program operation
The week program continues with the values corresponding to the current time.
Performance after power failure during sterilization operation
The sterilization process is cancelled and the controller changes to Fixed value operation mode with the
setpoints already entered. All setpoint values remain in memory.