22.4 Replacing the CO
2
sensor
Pull the power plug.
CAUTION
Risk of injuries by sharp edges on the rear panel.
Cuts.
Wear protective gloves when removing or installing the cover on the chamber rear
Remove the cover on the chamber rear
The position of the CO
2
sensor behind
the cover is shown in Figure 41.
Unscrew the cable from the sensor.
Make sure that the sensor cable does not fall
into the inner chamber.
Remove the fixing screws on both sides of the
sensor.
Pull the sensor backwards out of the holder. The
filter (white) attached to the sensor is pulled out
together with the sensor.
Then mount the new sensor proceeding in reverse order
Reinstall the cover on the chamber rear.
The sensor may only be used with the attached filter.
After replacing the CO
2
sensor repeat the CO
2
calibration.